Coating Thickness Gages TT260

Portable Coating Thickness Gauges TT260. New and improved model!

Features of TT260

Model TT260
Coating Thickness Gages TT260
  • Measuring principle: Eddy current & magnetic induction
  • Measuring modes: Single/continuous
  • Memory modes: Direct/batch
  • Onboard memory: 495 ( 5 batches of 99 readings each)
  • Stores up to 99 readings in Direct mode
  • Automatic calculation: Mean/Max/Min/No./S.DEV
  • Error detect ability: Buzz & display
  • Removable micro printer
  • Automatic shut-off
  • Operating temperature: 0??-40??
  • Power source: 5 x 1.2V (nickel-hydride batteries) User can replace battery
  • Weight: 1.4 lb (650g)
  • Dimensions: 270 x 86 x 47mm

Measuring Materials of TT260

Probe Model Probe Type Material of Coatings Material of substrates
F1, F10, F400 Ferrous Non-magnetic Magnetic
N1 Non-ferrous Insulating Non-magnetic metal
CN02 Non-ferrous Non-magnetic metal Insulating

??

Standard Accessories

  • 1 Calibration Certificate
  • 1 Main Unit with Removable Micro Printer
  • 1 Roll of Printing Paper
  • 1 Probe F1
  • 1 set of Test Plates with
  • 1 Steel Substrate
  • 1 Charger
  • 1 Instruction Manual
  • 1 Carrying Case

Optional Accessories

  • Probe F10
  • Probe F400
  • Probe CN02
  • Communication cable and software

Coating Thickness Gauge Specifications

Probemodel Measure range (µm) Resolution (µm) Tolerance (µm) Min curvature radius (mm) Min measure area (mm) Critical substrate thickness (mm)
1 point calibration 2 point calibration convex concave
F1 0-1250 1 ??3%H+1) ??(1~3)%H+1] 1.5 9 ??7 0.5
F10 0-10000 10 ??3%H+10) ??(1~3)%H+10] 10 30 ??40 2
F400 0-400 0.1 ??3%H+1) ??(1~3)%H+0.7] 1 5 ??3 0.2
N1 0-1250 1 ??3%H+1.5) ??(1~3)%H+1.5] 3 10 ??5 0.3
CN02 0-5000 1 ??3%H+1)
??
flat 3.5 No limit
??

Note: H-nominal value
Probe model F1 and F400 calibrate through unknown coating: Tolerance ??(3~10)%H+1] µm


 

时代仪器咨询专线:上海:021-60491398 13818683556 睢(sui)先生

|